Reachable Accuracy and Precision for Tritium Measurements by Calorimetry at TLK

C. G. Alecu, U. Besserer, B. Bornschein, B. Kloppe, Z. Köllö, J. Wendel

Fusion Science and Technology / Volume 60 / Number 3 / October 2011 / Pages 937-940

Measurement, Monitoring, and Accountancy / Proceedings of the Ninth International Conference on Tritium Science and Technology

At Tritium Laboratory Karlsruhe (TLK) calorimetry is used as the primary method for tritium accountability and for samples ranging from micrograms to tens of grams. An accuracy of 0.3% is achievable for samples containing more than 0.5 mg tritium and the precision of each calorimeter is better than 0.5%. The results of the TLK calorimeters vary within maximum 0.6% from each other.

 